Back to top
Send SMS Send Inquiry
MICROPROCESSOR FLAME PHOTOMETER 1385 & 1382

MICROPROCESSOR FLAME PHOTOMETER 1385 & 1382

Product Details:

X

Product Description

FEATURES

·         Microprocessor Based with Printer Interface

·         Auto Gas Cutoff (Model - 1385)

·         Auto Flame Failure Detection (Model - 1385)

·         Four Measurements in Single Aspiration

·         Setup Storage Facility

·         Sample Data Storage Facility

·         Curve Calibration Programmability Using Maximum of 5 Standards

·         Multiple Calibration Curve Saving Facility

·         Data Processing Through Curve Fitting Techniques, Least Square, Quadratic

·         Online Help Key

·         Direct Results in ppm and meq

·         Useful for Medical & Soil Testing

 

Microprocessor Flame Photometer - 1385 comes with flame failure detection and auto gas cutoff facility. In case of power failure, the gas cuts off automatically. In case the flame goes off accidently during work, the flame failure alarm sounds after few seconds 

Microprocessor Flame Photometers - 1385 & 1382 are used for the determination of sodium, potassium, calcium and lithium. They uses the latest microcontroller technology and advanced engineering techniques so as to give enhanced and reproducibility. They have soft touch membrane key for ease of operations.

The sample solution is aspirated through an automiser. Sample, Air and the fuel are mixed in the mixing chamber which is then sprayed as a very fine mist into the flame. The color of the flame is changed depending upon the concentration of elements present. Radiations from the flame passes through the sensing system and specific narrow band interference filter which permits only the characteristic radiation to pass to the photo-detector. The output of the photo-detector is then processed by the microcontroller and the final results are displayed on the 20 x 4 line alphanumeric lcd with backlit.

In both models 5-Point calibration facility and non-linearity correction is provided through an automatic curve fitting software. Facility has been provided for re-standardization with a single standard, thereby, eliminating the need of the multiple standards once a calibration has been performed.

 

SPECIFICATIONS:-

Mode

General

Serum

Urine

Range

Na: 0 - 100 ppm
Upto 250 meq/l, 1:100 dil

K: 0 - 100 ppm
Upto 250 meq/l, 1:100 dil

Ca: 0 - 100 ppm
Upto 250 meq/l, 1:100 dil

Li: 0 - 100 ppm
Upto 250 meq/l, 1:100 dil

Na: 0 - 200
1:100 dil

K: 0 - 10
1:100 dil

-


Li: 0 - 2
1:5 dil

Na: 0 - 250
1:100 dil

K: 0 - 100
1:100 dil

Ca: 0 - 10
1:2 dil

-

Sensitivity

General: Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm

Filters

Na & K supplied (Ca & Li are optional at extra cost)

Resolution

0.1 ppm/meq

Reproducibility

< 1% coefficient of variation for 20 consecutive samples using 10 ppm Na set as maximum standard

Curve Fit Accuracy

< 2% error when 3 ppm Na/K and 5 ppm Li are set as maximum standards

Display

20 x 4 alphanumeric LCD with backlit

Average Time

In-built in software

Flame System

LPG and dry oil free air

Detector

Silicon photodiode

Calibration

Upto 5-Point calibration with curve fitting software

Gas Control

Adjustable with knobs

Flame Failure

Auto detection (Model - 1385)

Gas Cuttoff

Automatic (Model - 1385)

Ignition System

Auto ignition

Automiser

Axial flow type

Printer

Provision for attachment of dot matrix printer or HP laserjet/ inkjet (non-windows) printers with centronics parallel interface

RS-232C

Available (PC software at extra cost)

Power

230 V + 10% AC, 50 Hz

Dimensions

285 x 255 x 210 mm (L x B x H) (Approx.)

Weight

7.5 Kg (Approx.)

COMPRESSOR UNIT

Air Supply

By oil free mini compressor unit with pressure regulator

Combustion Gas

LPG controlled by precision regulator

Power

230 V + 10% AC, 50 Hz

Dimensions

290 x 255 x 210 mm (L x B x H) (Approx.)

Weight

8 Kg (Approx.)

Enter Buying Requirement Details
Email Id
Mobile number



GST : 06ACOPS0805K1ZD
No-5309/27, Punjabi Mohalla, Ambala Cantt - 133001, Haryana, India
Phone :08068821194
Mr Devender Singhla (Chief Executive)
Mobile :08068821194